Photometer 10

Photometer 10

Handy single-beam filter photometer with discrete filters

TECHNICAL SPECIFICATIONS

Type Single-beam filter photometer
Light Source LED
Wavelength 450 nm – 650 nm
Wavelength Selection Single externally exchangeable interference filter, e.g. 546 nm
Photometric Range 0 – 2.5 A
0.5 – 4.0 McFarland units
Please note: Photometer 10 is also available as a densitometer for turbidity measurements of e.g. bacterial suspensions shown in McFarland units at 623 nm.
Cuvette System Normal standard cuvette (macro or semimicro (with measuring volume min 350μl), disposable or special optical glass)
or round cuvette (14 mm diameter) with measuring volume min 1500μl
Operator Interface Touchscreen for direct functions and alphanumerical inputs
Data Presentation Graphic display: white characters or symbols, blue background, lighted, resolution 128 x 64 dots
Languages English/ French/ German and Spanish
Memory
  • General operating software can be updated by PC
  • Data import by touchscreen or PC
Signal Port
  • USB serial port for connection to a PC
  • RS232 connector for connecting a barcode reader for an easy management of sample identities
Data Logging Up to 1000 results can be saved in the memory automatically
Measurement Procedures
  • Absorbance
  • Turbidity
Power Supply Input: 100 VAC up to 240 VAC at 50/60 Hz, output: 12 VDC / 1 ADC
or 4 x AA batteries
Weight 0.5 kg


Manuals available in:


Flyer available in: